Technical Guide for Endoscope Module Selection
I. Core Parameter Compatibility Analysis
1. Size Matching Assessment
Preferred Applications: Medical lumens (e.g., ureters, paranasal sinuses, periodontal pockets) and industrial conduits (Φ4.0mm+) with target channel diameters of 3.8–5.0mm
Critical Considerations: Evaluate sheath friction resistance for channel diameters 3.5–3.8mm
Inapplicable Scenarios: Flexible applications requiring channel diameters <3.3mm or large-angle bending (>90°)
2. Optical Performance Matching Analysis
Field of View Requirements: 88° FOV suits panoramic cavity observation; pair with optical zoom systems for localized magnification inspection
Working Distance Adaptation: 10-40mm fixed focal length suits close-range observation; select long-focus version for >50mm working distance
Illumination Uniformity: 6-LED ring layout suits 4-8mm diameter channels; <3mm channels may exhibit overexposure
II. Industry Application Adaptation Recommendations
Medical Field Selection Criteria
Sterilization Compatibility: Confirm high-temperature/high-pressure resistance of sheath material for reusable applications
Biocompatibility: Applications involving direct tissue contact must comply with ISO 10993 standards
Operational Ease: 25mm length suits straight endoscopes; flexible applications require bending radius assessment
Key Selection Criteria for Industrial Applications
Environmental Tolerance: Oil-contaminated environments require confirmation of front window lens anti-oil coating
Mechanical Strength: Pipeline inspection requires assessment of front-end impact resistance
Temperature Adaptability: Select temperature-resistant models for high-temperature environments (>60°C).
III. System Integration Technical Requirements
Electrical Interface Compatibility
Recommended image processor support: Minimum 720P@30fps real-time processing.
Power Supply System must provide stable LED drive current (constant current drive recommended).
Signal transmission cable lengths exceeding 1.5m require signal amplification solutions.
Mechanical Integration Solutions
Fixed installations should use M2.5 or larger threaded fasteners
Articulated joint integration requires ±0.3mm positional tolerance allowance
Waterproof applications require O-ring seals (silicone recommended) at connection points
IV. Cost-Effectiveness Optimization Recommendations
Scenarios for downgraded configurations
Routine inspections requiring only standard-definition imaging (480P)
Environments with fixed working distances and adequate illumination
Open inspections with non-critical dimensional requirements
Scenarios Requiring Upgraded Configuration
Applications requiring image measurement or 3D reconstruction
Extreme low-light environments (<10 lux)
Medical applications requiring frequent high-temperature sterilization
V. Validation Testing Recommendations
Mandatory Test Items
Imaging clarity testing under actual working conditions
Stability testing for continuous operation exceeding 4 hours
Performance testing under extreme temperature conditions
Recommended Test Items
Color reproduction accuracy on different tissue/material surfaces
Imaging quality in liquid media
Image stability under vibration
VI. Alternative Solutions
If this module does not fully meet requirements, consider:
Smaller diameter: 2.8mm version (field of view reduced to 75°)
Higher resolution: 2-megapixel version (size increased to 4.0mm)
Zoom version: Supports 15-60mm motorized focus